Z-0607-2018Class II
Synvisc-One (hylan G-F 20), 6mL delivered from a 10mL syringe
Sanofi Genzyme Corp. / initiated 2017-12-11 / Terminated / root cause: manufacturing process
Sanofi Genzyme began a recall of Synvisc-One (hylan G-F 20), 6mL delivered from a 10mL syringe Product Usage: A elastoviscous high molecular weight fluid containing hylan A and hylan B... on . FDA classified it as Class II and gives the reason as "Contamination with Methylbacterium thiocyanatum". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0607-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sanofi Genzyme Corp.
- Product
- Synvisc-One (hylan G-F 20), 6mL delivered from a 10mL syringe Product Usage: A elastoviscous high molecular weight fluid containing hylan A and hylan B polymers produced from chicken combs. Synvisc-One is administered as a single intra-articular injection. Synvisc-One is indicated for the treatment of pain in osteoarthritis (OA) of the knee in patients who have failed to respond adequately to conservative non-pharmacologic therapy and simple analgesics, e.g., acetaminophen.
- Product code
- MOZ Acid, Hyaluronic, Intraarticular
- Reason for recall
- Contamination with Methylbacterium thiocyanatum
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- On December 11, 2017 Sanofi Genzyme distributed URGENT VOLUNTARY DEVICE RECALL notices and reply cards to their customers via courier service. Customers were advised to inspect inventory, immediately discontinue use and return all affected product. On December 19, 2017 Sanofi Genzyme distributed a 2nd notice titled DEVICE RECALL UPDATE NOTICE (2 of 3) notifying customers that the isolated bacterial organism was Methylbacterium thiocyanatum. On December 22, 2017 Sanofi Genzyme distributed a 3rd notice titled DEVICE RECALL UPDATE NOTICE (3 of 3) which included antibiotic sensitivity analysis guidance for the isolated organism. Customers who wish to return product should contact GENCO Pharmaceutical Services at 1-855-838-5782 for a recalled product return kit. Customers are encouraged to complete and return the Business Reply Card via fax to 1-414-459-8791 or via email to [email protected]. Sanofi Genzyme will issue you a credit or replacement product upon receipt of the recalled product by GENCO. All questions of a medical or clinical nature should be directed to Sanofi Medical Information Services in the U.S. at 1-800-633-1610, option 1. Sanofi US Customer service can be reached at 1-800-633-1610 option 7 then 4.
- Quantity in commerce
- 12,380 units
- Distribution
- US Nationwide distribution(36 states)
- Product codes and lots
- Lot Number: 7RSL021 Expiration Date: 2020-05-31