Z-0392-2018Class II

BD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for collection and transport of vaginal...

Becton Dickinson & Co / initiated 2017-07-18 / Terminated

Becton Dickinson & Co. began a recall of BD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for collection and transport of vaginal and endocervical swab specimens, in addition to... on . FDA classified it as Class II and gives the reason as "BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0392-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.
Product code
NSU Instrumentation For Clinical Multiplex Test Systems
Reason for recall
BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead to repeat of the test and/or re-collection of a specimen.
FDA root cause
Use error
Action
Customers were notified via UPS on 07/18/2017. The recall for BD MAX" Vaginal Panel and BD MAX" UVE Specimen Collection Kit was initiated on July 18, 2017 and expanded to include BD MAX" CT/GC/TV on August 11, 2017. The recall expansion letter to customers was dated 08/14/2017.
Distribution
US and OUS.
Product codes and lots
UDI: 30382904433766. All Lot numbers: Not Distributed in US Not Distributed in US

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0392-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0392-2018Class IITerminatedVoluntary: Firm initiated
Z-0393-2018Class IITerminatedVoluntary: Firm initiated
Z-0394-2018Class IITerminatedVoluntary: Firm initiated