Vado Steerable Sheath 8.8F, REF SS8FMU67, packaged in a lidded tray that is within a pouch which is inserted...
Kalila Medical began a recall of Vado Steerable Sheath 8.8F, REF SS8FMU67, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on... on . FDA classified it as Class II and gives the reason as "Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0563-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Kalila Medical, Inc.
- Product
- Vado Steerable Sheath 8.8F, REF SS8FMU67, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- The recalling firm issued a letter dated 12/14/2017 via FedEx on 12/14/2017.
- Quantity in commerce
- 99 sheaths
- Distribution
- Distribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution.
- Product codes and lots
- Lot #0559, exp. 10/9/2018, and Lot #0571, exp. 10/19/2018.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0563-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0564-2018 | Class II | Terminated | Voluntary: Firm initiated |