Z-0564-2018Class II

Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch...

Kalila Medical, Inc. / initiated 2017-12-14 / Terminated / root cause: manufacturing process

Kalila Medical began a recall of Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx... on . FDA classified it as Class II and gives the reason as "Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0564-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Kalila Medical, Inc.
Product
Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.
Product code
DYB Introducer, Catheter
Reason for recall
Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
The recalling firm issued a letter dated 12/14/2017 via FedEx on 12/14/2017.
Quantity in commerce
261 sheaths
Distribution
Distribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution.
Product codes and lots
Lot #0594, exp. 1/19/2018, and Lot #0626, exp. 04/23/2018.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0564-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0563-2018Class IITerminatedVoluntary: Firm initiated
Z-0564-2018Class IITerminatedVoluntary: Firm initiated