smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating...
Smith & Nephew, Inc. began a recall of smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD... on . FDA classified it as Class II and gives the reason as "Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0396-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442024, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa
- Root cause tag
- labeling and UDI
- FDA root cause
- Mixed-up of materials/components
- Action
- The firm, Smith & Nephew, Inc., sent an "Urgent Medical Device Recall Notice" letter dated December 11, 2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect your inventory, locate unused devices and quarantine immediately; if you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out; return product, and complete and return the response form by email to [email protected] or fax to 901-566-7975 even if you have not product to return. If you have any questions or concerns regarding this recall please contact [email protected] or call 901-566-7084.
- Quantity in commerce
- 2 units
- Distribution
- US Distribution to states of: MI, GA, and OH.
- Product codes and lots
- Lot number 15GM08498C
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0396-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0397-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.