Z-0402-2018Class II

Brilliance iCT SP - Model 728311 Brilliance iCT - Model 728306 Computed Tomography X-ray systems

Philips North America LLC / initiated 2017-07-11 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance iCT SP - Model 728311 Brilliance iCT - Model 728306 Computed Tomography X-ray systems on . FDA classified it as Class II and gives the reason as "Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0402-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance iCT SP - Model 728311 Brilliance iCT - Model 728306 Computed Tomography X-ray systems
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.
Root cause tag
software
FDA root cause
Software design
Action
A Customer Information Letter dated June 29, 2017, was sent to US consignees via certified mail. Philips Key Market representatives will distribute the letter internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
Quantity in commerce
559 total
Distribution
Worldwide distribution, including US nationwide.
Product codes and lots
Brilliance iCT 4.1.6 version. Brilliance iCT SP - Model 728311 Codes: 200038 200203 200119 200052 200206 200210 200208 200213 200097 200100 200063 200009 200011 200204 200212 200018 200207 200027 200130 83001 200209 200215 200043 200032 200200 200051 200086 200202 200211 200024 200023 200068 200201 200017 200061 200205 200030 200117 200044 200015 83000 200022 200093 200101 200040 200013 200107 200111 200138 200214 200102 200047 200019 200105 200109 200121 200041 200131 200127 200003 200092 200074 200088 200067 200089 200049

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0402-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0402-2018Class IITerminatedVoluntary: Firm initiated
Z-0403-2018Class IITerminatedVoluntary: Firm initiated