Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable...
Philips Electronics North America Corporation began a recall of Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It... on . FDA classified it as Class II and gives the reason as "If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0311-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
- FDA root cause
- Under Investigation by firm
- Action
- If the lock mechanism of the movable clamp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is . applied. This may happen when the detector is rotated clockwise from landscape to portrait position. The Wireless Portable Detector (WPD) is fixed in the Mobile Detector Holder via a lock mechanisrn. The 4 countersunk screws, that are used to attach the lock mechanism, can become loose. If this happens and the mechanism fails, the detector may fall down when the holder is rotated 90 clockwise from Iandscape.
- Quantity in commerce
- 412
- Distribution
- 412 units affected.
- Product codes and lots
- 9890-010-88152 - Codes included in attached list. Devices were manufactured and distributed from March 2015 - March 2017.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0311-2018 | Class II | Terminated | Voluntary: Firm initiated |