Z-0311-2018Class II

Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable...

Philips North America LLC / initiated 2017-07-14 / Terminated

Philips Electronics North America Corporation began a recall of Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It... on . FDA classified it as Class II and gives the reason as "If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0311-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.
Product code
KPR System, X-Ray, Stationary
Reason for recall
If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
FDA root cause
Under Investigation by firm
Action
If the lock mechanism of the movable clamp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is . applied. This may happen when the detector is rotated clockwise from landscape to portrait position. The Wireless Portable Detector (WPD) is fixed in the Mobile Detector Holder via a lock mechanisrn. The 4 countersunk screws, that are used to attach the lock mechanism, can become loose. If this happens and the mechanism fails, the detector may fall down when the holder is rotated 90 clockwise from Iandscape.
Quantity in commerce
412
Distribution
412 units affected.
Product codes and lots
9890-010-88152 - Codes included in attached list. Devices were manufactured and distributed from March 2015 - March 2017.