Z-0566-2018Class II
Medtronic ANGLED DRILL, REF 6630902, QTY: 1 EA Rx only
Medtronic Sofamor Danek USA Inc / initiated 2017-12-22 / Terminated / root cause: manufacturing process
Medtronic Sofamor Danek USA Inc began a recall of Medtronic ANGLED DRILL, REF 6630902, QTY: 1 EA Rx only on . FDA classified it as Class II and gives the reason as "Medtronic has discovered the product lot in question was noted as having a longer drill tip extension than what is indicated on the print specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0566-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- Medtronic ANGLED DRILL, REF 6630902, QTY: 1 EA Rx only
- Product code
- MDM Instrument, Manual, Surgical, General Use
- Reason for recall
- Medtronic has discovered the product lot in question was noted as having a longer drill tip extension than what is indicated on the print specification.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated the recall by letter on 12/22/2017, requesting the return of the product.
- Quantity in commerce
- 2 units
- Distribution
- MN, NC
- Product codes and lots
- UDI: 00643169363502, Lot Number EM17F015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0566-2018 | Class II | Terminated | Voluntary: Firm initiated |