CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general...
Philips Electronics North America Corporation began a recall of CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine... on . FDA classified it as Class II and gives the reason as "The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0330-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Philips Medical System sent an Customer Notification letter dated October 19, 2017, and plans to have a Philips Field Service Engineer go to site to perform the field change order (FCO70900041) including: 1. Check for the presence of the PE cable and, if missing, install it. 2 .Perform a modified screw fixation. They expect to complete action by May 2018. For further questions, please call (978) 659-3000.
- Quantity in commerce
- 10
- Distribution
- Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.
- Product codes and lots
- All CombiDiagnost R90 UDI 709030
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0330-2018 | Class II | Terminated | Voluntary: Firm initiated |