Z-0330-2018Class II

CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general...

Philips North America LLC / initiated 2017-10-19 / Terminated / root cause: manufacturing process

Philips Electronics North America Corporation began a recall of CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine... on . FDA classified it as Class II and gives the reason as "The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0330-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Philips Medical System sent an Customer Notification letter dated October 19, 2017, and plans to have a Philips Field Service Engineer go to site to perform the field change order (FCO70900041) including: 1. Check for the presence of the PE cable and, if missing, install it. 2 .Perform a modified screw fixation. They expect to complete action by May 2018. For further questions, please call (978) 659-3000.
Quantity in commerce
10
Distribution
Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.
Product codes and lots
All CombiDiagnost R90 UDI 709030

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0330-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0330-2018Class IITerminatedVoluntary: Firm initiated