MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
Stryker Medical Division of Stryker Corporation began a recall of MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM on . FDA classified it as Class II and gives the reason as "It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0419-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Medical
- Product
- MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Stryker notified customers of the recall by sending Customer Notification and Business Reply Forms, dated July 21, 2017.
- Quantity in commerce
- 11160
- Distribution
- U.S.
- Product codes and lots
- LOT #(s): 1640002, 1645003
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0419-2018 | Class II | Terminated | Voluntary: Firm initiated |