Z-0419-2018Class II

MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM

Stryker Medical / initiated 2017-07-21 / Terminated / root cause: design

Stryker Medical Division of Stryker Corporation began a recall of MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM on . FDA classified it as Class II and gives the reason as "It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0419-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Medical
Product
MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
Product code
DWJ System, Thermal Regulating
Reason for recall
It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).
Root cause tag
design
FDA root cause
Component design/selection
Action
Stryker notified customers of the recall by sending Customer Notification and Business Reply Forms, dated July 21, 2017.
Quantity in commerce
11160
Distribution
U.S.
Product codes and lots
LOT #(s):  1640002, 1645003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0419-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0419-2018Class IITerminatedVoluntary: Firm initiated