Z-0385-2018Class II

Aspire HD Plus (FDR MS-2500), also known as Aspire HD Plus F and Aspire HD-s (FDR MS-2000), also known as...

FUJIFILM Healthcare Americas Corporation / initiated 2017-07-14 / Terminated / root cause: manufacturing process

Fujifilm Medical Systems U.S.A., Inc. began a recall of Aspire HD Plus (FDR MS-2500), also known as Aspire HD Plus F and Aspire HD-s (FDR MS-2000), also known as Aspire HD Plus S). Mammography machine The Fujifilm... on . FDA classified it as Class II and gives the reason as "There was a case that the swivel arm went down unintentionally when it was moved up/down". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0385-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Aspire HD Plus (FDR MS-2500), also known as Aspire HD Plus F and Aspire HD-s (FDR MS-2000), also known as Aspire HD Plus S). Mammography machine The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other fall-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
There was a case that the swivel arm went down unintentionally when it was moved up/down.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
FUJIFILM sent an Urgent: Medical Device Correction Letter dated July 14, 2017, accompanied by a Customer Feedback form to customers. PLEASE READ AND FOLLOW THE INSTRUCTIONS Please refer to the following page which provides the details of the problem and instructions for actions to be taken. Please follow the instructions in the ACTIONS TO BE TAKEN BY CUSTOMER/USER section. We sincerely regret the inconvenience that this may cause you. FUJIFILM is committed to providing products and services of the highest quality. Your satisfaction with FUJIFILM products and with our response to this issue is very important to us. If you have any questions about this matter, please contact your local FUJIFILM office. For further questions, please call (360) 356-6821.
Quantity in commerce
20 US
Distribution
US Distribution and Internationally to Mexico.
Product codes and lots
FDR MS-2500 made in Japan : S/N- 16640001 or later  FDR MS-2500 made in China : S/N- 26860001 or later  FDR MS-2000 made in Japan : S/N- 16650001 or later  FDR MS-2000 made in China : S/N- 26870001 or later

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0385-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0385-2018Class IITerminatedVoluntary: Firm initiated
Z-0386-2018Class IITerminatedVoluntary: Firm initiated