Philips Healthcare Brilliance iCT SP computed tomography x-ray system
Philips Medical Systems (Cleveland) Inc began a recall of Philips Healthcare Brilliance iCT SP computed tomography x-ray system on . FDA classified it as Class II and gives the reason as "There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0332-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Healthcare Brilliance iCT SP computed tomography x-ray system
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips Healthcare is informing consignees of this issue through a Customer Information Letter (CIL_FCO72800690). Philips Key Market representatives will distribute the letter internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order. Customers are notified that if they hear an unusual noise coming from the gantry, or scan halts, the Philips representative should be contacted as soon as possible. Philips will be inspecting the affected systems, tightening and replacing the reaction ring screws as needed at no additional cost.
- Quantity in commerce
- 134
- Distribution
- 107 domestic accounts 15 US Govt accounts 682 Foreign Accounts 3 Mexico accounts 12 Canada accounts
- Product codes and lots
- Brilliance iCT SP - 728311 Devices manufactured 03/05/2009 to present.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0331-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0332-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0333-2018 | Class II | Terminated | Voluntary: Firm initiated |