Z-0332-2018Class II

Philips Healthcare Brilliance iCT SP computed tomography x-ray system

Philips North America LLC / initiated 2017-11-08 / Terminated / root cause: design

Philips Medical Systems (Cleveland) Inc began a recall of Philips Healthcare Brilliance iCT SP computed tomography x-ray system on . FDA classified it as Class II and gives the reason as "There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0332-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare Brilliance iCT SP computed tomography x-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.
Root cause tag
design
FDA root cause
Device Design
Action
Philips Healthcare is informing consignees of this issue through a Customer Information Letter (CIL_FCO72800690). Philips Key Market representatives will distribute the letter internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order. Customers are notified that if they hear an unusual noise coming from the gantry, or scan halts, the Philips representative should be contacted as soon as possible. Philips will be inspecting the affected systems, tightening and replacing the reaction ring screws as needed at no additional cost.
Quantity in commerce
134
Distribution
107 domestic accounts 15 US Govt accounts 682 Foreign Accounts 3 Mexico accounts 12 Canada accounts
Product codes and lots
Brilliance iCT SP - 728311  Devices manufactured 03/05/2009 to present.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0332-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0331-2018Class IITerminatedVoluntary: Firm initiated
Z-0332-2018Class IITerminatedVoluntary: Firm initiated
Z-0333-2018Class IITerminatedVoluntary: Firm initiated