RayStation
RAYSEARCH LABORATORIES AB began a recall of RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit... on . FDA classified it as Class II and gives the reason as "For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0380-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Ray Search sent a Field Safety Notice letter dated July 14, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter included a response form with instructions for workarounds until the software is corrected.
- Quantity in commerce
- 117 units
- Distribution
- US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
- Product codes and lots
- 6.0.0.24 , 6.1.0.26 , 6.1.1.2 or 6.2.0.7
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0380-2018 | Class II | Terminated | Voluntary: Firm initiated |