Z-0380-2018Class II

RayStation

RaySearch Laboratories AB (publ) / initiated 2017-07-17 / Terminated / root cause: design

RAYSEARCH LABORATORIES AB began a recall of RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit... on . FDA classified it as Class II and gives the reason as "For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0380-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.
Root cause tag
design
FDA root cause
Device Design
Action
Ray Search sent a Field Safety Notice letter dated July 14, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter included a response form with instructions for workarounds until the software is corrected.
Quantity in commerce
117 units
Distribution
US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
Product codes and lots
6.0.0.24 , 6.1.0.26 , 6.1.1.2 or 6.2.0.7

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0380-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0380-2018Class IITerminatedVoluntary: Firm initiated