Z-1180-2018Class II

BD Vacutainer Barricor" Lithium Heparin Plasma Blood Collection Tubes for Invitro Diagnostic Use, Catalog...

Becton Dickinson & Co / initiated 2017-09-18 / Terminated / root cause: manufacturing process

Becton Dickinson & Company began a recall of BD Vacutainer Barricor" Lithium Heparin Plasma Blood Collection Tubes for Invitro Diagnostic Use, Catalog Numbers: 365043, 365044, 365045 & 365047 Used to... on . FDA classified it as Class II and gives the reason as "Increased amount of residual blood present on the top of stopper well after venipuncture". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1180-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Vacutainer Barricor" Lithium Heparin Plasma Blood Collection Tubes for Invitro Diagnostic Use, Catalog Numbers: 365043, 365044, 365045 & 365047 Used to collect, separate, process, transport and store venous blood samples for use in chemistry determinations, therapeutic drug monitoring and zinc testing in plasma for in vitro diagnostic use.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
Increased amount of residual blood present on the top of stopper well after venipuncture
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On September 22, 2017, BD distributed URGENT PRODUCT REMOVAL RECALL notices to distributors via email. Distributors and customers were mailed via USPS. Customers are advised to take to take the following actions: 1. Discontinue use of the tubes. 2. Complete and return the Customer Recall Response Form via fax to 1-855-544-4851 or via email to [email protected]. 3. Return all affected products using the enclosed packaging instructions and include the completed Recall Response Form with the product. Upon receipt of the returned product, BD will issue alternate product. Customers with questions may call 1-855-215-5168, Monday through Friday between 8AM and 5 PM, EST. For further questions, please call (201) 847-6800.
Quantity in commerce
1,976,000 tubes
Distribution
USA (nationwide)
Product codes and lots
All Lot Numbers Expiration Dates: November 30, 2017 - July 31, 2018  UDI No.: 50382903650434 50382903650441 50382903650458 50382903650472

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1180-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1180-2018Class IITerminatedVoluntary: Firm initiated