Brilliance 64, (Code 728231) Intended to produce images of the head and body by computer reconstruction of...
Philips Medical Systems (Cleveland) Inc began a recall of Brilliance 64, (Code 728231) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles... on . FDA classified it as Class II and gives the reason as "Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0389-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Brilliance 64, (Code 728231) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
- FDA root cause
- Under Investigation by firm
- Action
- An Urgent - Field Safety Notice Medical Device Correction letter, dated November 1, 2017, was sent to customers. The letter identified the affected device, described the issue, and discussed the hazard involved. Users are to cease use immediately and contact their Philips Field Service Engineer, if any unusual noises are heard within the gantry, or if the patient support tabletop is in free float. Philips Field Service Engineers will correct the system by replacing the affected fasteners. If further info or support concerning this issue is needed, customers can contact their local Philips rep or the Customer Care Solutions Center at 1-800-722-9377.
- Quantity in commerce
- 1
- Distribution
- Worldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts
- Product codes and lots
- Code 728231, SN # 10655
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0388-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0389-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0390-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0391-2018 | Class II | Terminated | Voluntary: Firm initiated |