Z-0705-2018Class II
Rusch Greenlite MAC 1, Catalog Number 004551001. Used during tracheal intubation
Teleflex Medical / initiated 2017-01-10 / Terminated / root cause: packaging
Teleflex Medical began a recall of Rusch Greenlite MAC 1, Catalog Number 004551001. Used during tracheal intubation on . FDA classified it as Class II and gives the reason as "The pouch may contain the incorrect Rusch Green Lite product size. Pouches that are labeled as containing RUSCH GREEN LITE MAC 1 may instead contain RUSCH GREEN LITE MAC 2". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0705-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Rusch Greenlite MAC 1, Catalog Number 004551001. Used during tracheal intubation.
- Product code
- CCW Laryngoscope, Rigid
- Reason for recall
- The pouch may contain the incorrect Rusch Green Lite product size. Pouches that are labeled as containing RUSCH GREEN LITE MAC 1 may instead contain RUSCH GREEN LITE MAC 2.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- The firm initiated their recall by letter on 01/10/2018 via FedEx 2-day. Affected consignees were instructed to take the following actions: "1. If you have affected stock, immediately discontinue use and quarantine any products with the product code and lot number listed above, so that the affected products can be returned to Teleflex. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter."
- Quantity in commerce
- 2140 units
- Distribution
- US only
- Product codes and lots
- Lot/Serial number: 1602312
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0705-2018 | Class II | Terminated | Voluntary: Firm initiated |