Z-0638-2018Class II
ROSA Brain 3.0 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning...
Zimmer GmbH / initiated 2017-05-01 / Terminated / root cause: software
Zimmer Biomet, Inc. began a recall of ROSA Brain 3.0 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning and orientation of instrument holders to be used... on . FDA classified it as Class II and gives the reason as "Robot arm being sent to the wrong position". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0638-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ROSA Brain 3.0 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning and orientation of instrument holders to be used by neurosurgeons.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Robot arm being sent to the wrong position
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Urgent Medical Device Recall (Correction) notices were sent to customers. The letter provided to hospital risk managers and surgeons identified the issue and their responsibilities. These responsibilities included: - Stop using the micromovement function in isocentric mode during Brain Surgery with ROSA Brain 3.0 device. -Reviewing the notification to ensure comprehension of the steps required to be taken prior to the permanent correction that will be implemented on site by a service engineer. - Completing and returning Attachment 1 (Certificate of Acknowledgement) to the e-mail address provided. The firm has prepared a software correction that will be deployed and implemented on site by a MEDTECH representative.
- Quantity in commerce
- 16
- Distribution
- Distributed to CA, DC, FL, IL, KY, MA, MN, NC, NY, OH, and PA.
- Product codes and lots
- BR16005, BR16004, BR16006, BR16012, BR16011, BR16009, BR16010, BR16014, BR16013, BR16015, BR16021, BR16017, BR16018, BR16016, BR16023, and BR16025.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0638-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0639-2018 | Class II | Terminated | Voluntary: Firm initiated |