Z-0639-2018Class II

ROSA Spine 1.0.2 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial...

Zimmer GmbH / initiated 2017-05-01 / Terminated / root cause: software

Zimmer Biomet, Inc. began a recall of ROSA Spine 1.0.2 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning and orientation of instrument holders to be... on . FDA classified it as Class II and gives the reason as "Robot arm being sent to the wrong position". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0639-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Spine 1.0.2 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning and orientation of instrument holders to be used by surgeons.
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Robot arm being sent to the wrong position
Root cause tag
software
FDA root cause
Software design
Action
Urgent Medical Device Recall (Correction) notices were sent to customers. The letter provided to hospital risk managers and surgeons identified the issue and their responsibilities. These responsibilities included: - Stop using the micromovement function in isocentric mode during Brain Surgery with ROSA Brain 3.0 device. -Reviewing the notification to ensure comprehension of the steps required to be taken prior to the permanent correction that will be implemented on site by a service engineer. - Completing and returning Attachment 1 (Certificate of Acknowledgement) to the e-mail address provided. The firm has prepared a software correction that will be deployed and implemented on site by a MEDTECH representative.
Quantity in commerce
1
Distribution
Distributed to CA, DC, FL, IL, KY, MA, MN, NC, NY, OH, and PA.
Product codes and lots
SP16008

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0639-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0638-2018Class IITerminatedVoluntary: Firm initiated
Z-0639-2018Class IITerminatedVoluntary: Firm initiated