510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0604-2018Class II

11MM/130 Degree Titanium Trochanteric Fixation Nail 380MM/Right, Sterile

Synthes (USA) Products LLC / initiated 2017-07-21 / Terminated / root cause: manufacturing process

Synthes (USA) Products LLC began a recall of 11MM/130 Degree Titanium Trochanteric Fixation Nail 380MM/Right, Sterile on . FDA classified it as Class II and gives the reason as "The locking mechanism and protective cap were missing from the sterile packed nails of the affected lot". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0604-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
11MM/130 Degree Titanium Trochanteric Fixation Nail 380MM/Right, Sterile
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
The locking mechanism and protective cap were missing from the sterile packed nails of the affected lot.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The company issued a recall letter on 7/21/2017 asking customers to quarantine affected product and arrange for it to be returned.
Quantity in commerce
4 units
Distribution
TX, GA, PA, UT and Canada
Product codes and lots
Lot H302839, Expiration Date 31Jan2026

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0604-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0604-2018Class IITerminatedVoluntary: Firm initiated