510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0601-2018Class II

ARROW PICC powered by ARROW VPS Stylet, Product Codes: CDC-35052-VPS and CDC-35041-VPS. Peripherally Inserted...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2017-07-25 / Terminated / root cause: packaging

Arrow International Inc began a recall of ARROW PICC powered by ARROW VPS Stylet, Product Codes: CDC-35052-VPS and CDC-35041-VPS. Peripherally Inserted Central Catheter on . FDA classified it as Class II and gives the reason as "Finished good kits may contain incorrect components. Kits that should contain a 5 FR 2 lumen catheter may contain a 4 FR single lumen catheter and kits that should contain a 4 FR single lumen catheter may contain a 5 FR...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0601-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
ARROW PICC powered by ARROW VPS Stylet, Product Codes: CDC-35052-VPS and CDC-35041-VPS. Peripherally Inserted Central Catheter
Product code
LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Reason for recall
Finished good kits may contain incorrect components. Kits that should contain a 5 FR 2 lumen catheter may contain a 4 FR single lumen catheter and kits that should contain a 4 FR single lumen catheter may contain a 5 FR 2 lumen catheter. This may cause a delay in treatment while a replacement catheter is obtained.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Arrow International notified all U.S. consignees with letter sent on July 25, 2017, via FedEx 2-day. second notification August 22, 2017, third notification November 28, 2017.
Quantity in commerce
2691 total
Distribution
US nationwide distribution.
Product codes and lots
Lot/Batch Numbers: 23F17A0038, 23F17A0040, 23F16M0059.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0601-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0601-2018Class IITerminatedVoluntary: Firm initiated
Z-0602-2018Class IITerminatedVoluntary: Firm initiated