Z-0534-2018Class II

Covidien Vistec" X-Ray Detectable Sponges, 7317 Vistec" X-Ray Detectable Sponges Sponge

Medtronic, Inc. / initiated 2017-07-17 / Terminated / root cause: manufacturing process

Covidien LP began a recall of Covidien Vistec" X-Ray Detectable Sponges, 7317 Vistec" X-Ray Detectable Sponges Sponge. product Usage: Vistec X-Ray Detectable Sponges are medical absorbent... on . FDA classified it as Class II and gives the reason as "Medtronic is voluntarily recalling two production lots of its Covidien Vistec" x-ray detectable sponges de to possible contamination with human blood caused by handling of the product during the manufacturing process...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0534-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien Vistec" X-Ray Detectable Sponges, 7317 Vistec" X-Ray Detectable Sponges Sponge. product Usage: Vistec X-Ray Detectable Sponges are medical absorbent devices intended to be used internally within the body cavity to control bleeding or for applying medication. These sponges contain unique dual-colored elements that combines blue radiopaque strand with white strand, which is visible when sponge is saturated with blood.
Product code
GDY Gauze/Sponge, Internal, X-Ray Detectable
Reason for recall
Medtronic is voluntarily recalling two production lots of its Covidien Vistec" x-ray detectable sponges de to possible contamination with human blood caused by handling of the product during the manufacturing process and prior to final packaging and Ethylene Oxide sterilization.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Medtronic sent an Urgent Medical Device Recall letter dated July 17, 2017 to their customers via Federal Express or certified mail. The letter identified the affected product, problem and actions to be taken. Customers are requested to quarantine, discontinue use and return affected product. Customers were instructed to acknowledge receipt of the Urgent Field Corrective Action Notice. For questions contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Quantity in commerce
33,280
Distribution
U.S., Gov Accts, and foreign countries: PA, PR, CO.
Product codes and lots
Lot Numbers: 17E121362, 17E121262 UDI: 10884527016990; 20884527016997

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0534-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0534-2018Class IITerminatedVoluntary: Firm initiated