Z-0614-2018Class II

Artis oneFloor-mounted system for uncompromised imaging; Artis one, Interventional fluoroscopic x-ray system...

Siemens Medical Solutions USA, Inc. / initiated 2017-08-22 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, INC began a recall of Artis oneFloor-mounted system for uncompromised imaging; Artis one, Interventional fluoroscopic x-ray system, Artis one is an angiography system developed for... on . FDA classified it as Class II and gives the reason as "The possibility exists for the monitor display of Artis one systems with a DP Transceiver below revision 03 to experience blinking and/or a blue/black screen". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0614-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis oneFloor-mounted system for uncompromised imaging; Artis one, Interventional fluoroscopic x-ray system, Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. X-ray generator POLYDOROS A100 Plus
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
The possibility exists for the monitor display of Artis one systems with a DP Transceiver below revision 03 to experience blinking and/or a blue/black screen
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent- Medical Device Correction letter dated August 22, 2017. Our service organization will contact you shortly to arrange a date to perform this corrective action (A system hardware update will be implemented and all affected DP Transceivers below revision 03 will be updated). Following the system hardware update, the cause is eliminated and a recurrence of the fault is prevented. For further questions, please call (610) 448-6461.
Quantity in commerce
207 worldwide (13 distributed in US)
Distribution
US Distribution to the states of : AR, MO, NY, IN, AR, TX , NC, PA, MI, LA, WA, CA and AL.
Product codes and lots
Serial # 82044; 82105; 82070; 82054; 82040;82038;82065 82108; 82010; 82039; 82109 82102 82066

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0614-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0614-2018Class IITerminatedVoluntary: Firm initiated