Z-0535-2018Class II

Medtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary...

Medtronic Perfusion Systems / initiated 2018-01-03 / Terminated

Medtronic Perfusion Systems began a recall of Medtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures on . FDA classified it as Class II and gives the reason as "Medtronic custom perfusion tubing packs contain Terumo OPS valves that were recalled due to a lack of flow through the valve". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0535-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Perfusion Systems
Product
Medtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Medtronic custom perfusion tubing packs contain Terumo OPS valves that were recalled due to a lack of flow through the valve.
FDA root cause
Component change control
Action
In response to an Urgent Safety Alert issued by Terumo Cardiovascular Systems (CVS) Medtronic initiated their safety alert verbally to their direct consignees on 10/23/2017, providing them with a copy of the Terumo Safety Alert dated 06/27/2017. The firm followed this action with a letter on 1/3/18 in which they stated that the action by Terumo CVS was changed to a recall, and , in response, Medtronic was following up as a recall and requested the products be returned.
Quantity in commerce
2914 units
Distribution
DE, IA, CA and Japan
Product codes and lots
a) REF 2815R1, UPN/GTIN: 20885074191038, Lot/Serial Numbers: 210937492, 213151103, 213671744  b) REF 2D25R6, UPN/GTIN: 00613994268174, Lot/Serial Numbers: 209533592, 209759280, 209948919, 210174899, 210377603,  210478517, 210591995, 210767149, 211011818, 211210404,  211243436, 211542842, 211761616  c) REF 9M68R, UPN/GTIN: 20643169876648, Lot/Serial Numbers: 212067087, 212796361  d) REF BB4B15R13, UPN/GTIN: 00643169845121, Lot/Serial Numbers: 211804184, 212073924, 212266962, 212268301, 212385983, 212552883,  212796215, 213248430, 213583108  e) REF BB4B15R14, UPN/GTIN: 00763000025878, Lot/Serial Numbers: 213801240, 213868837  f) REF BB9L72R, UPN/GTIN: 00643169845077, Lot/Serial Numbers: 211715103  g) REF BB9L72R1, UPN/GTIN: 00643169872127, Lot/Serial Numbers:  212020797, 212174623, 212193001, 212301586, 212355584, 212428238,  212626142, 212884370, 213001157, 213248459, 213631247, 213672934,  213802501, 214072742, 214093025  h) REF CB1Z87R11, UPN/GTIN: 00643169741591, Lot/Serial Numbers: 211755326  i) REF CB4L56R24, UPN/GTIN: 00643169735989, Lot/Serial Numbers: 211654744, 211654745, 211800954  j) REF CB4L56R26, UPN/GTIN: 00643169906181, Lot/Serial Numbers: 212294432, 212404544, 212447345, 212489670, 213251146, 213251166,  213358174, 213573495, 213623995, 214053395  k) REF CB5S68R3, UPN/GTIN: 20613994490760, Lot/Serial Numbers: 210937507  l) REF CB7X82R4, UPN/GTIN: 00643169471986, Lot/Serial Numbers: 211050766, 211320749  m) REF CB7X82R5, UPN/GTIN: 00643169789517, Lot/Serial Numbers:  211423503, 211443209, 211443210, 211885323, 211991293, 212154744,  212235237, 212274655, 212428312, 212473311, 212671632, 212962650,  213126447, 213292357, 213377926, 213445426, 213494394, 213494396,  213672914, 213906498, 214043035, 214093030  n) REF HY6Y19R5, UPN/GTIN: 00643169510432, Lot/Serial Numbers: 210802114, 210945613, 211114272, 211210180, 211327621, 211654742,  211846417, 212235012, 212545229, 212835348  o) REF HY8M65R2, UPN/GTIN: 00643169661295, Lot/Serial Numbers: 210952226, 211335076, 211558225, 211722651, 212028419, 212222066,  212355669, 213001001, 213719738  p) REF TL8W39R6, UPN/GTIN: 00643169715097, Lot/Serial Numbers: 210731697, 210838201, 210937704

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0535-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0535-2018Class IITerminatedVoluntary: Firm initiated