AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125...
Beckman Coulter Inc. began a recall of AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system on . FDA classified it as Class II and gives the reason as "A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1151-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A recall notification letter was sent to affected US consignees on 1/19/2018 notifying them of the issue. In addition, the recalling firm is instructing field service engineers to inspect the tip present sensor for misalignment at the next service visit.
- Quantity in commerce
- 73 units in US
- Distribution
- Worldwide distribution. US nationwide, Algeria, Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
- Product codes and lots
- All units in the AutoMate 2500 Family are affected; Serial Numbers - 1107, 4048, 4055, 4025, 4052, 4079, 4091, 4076, 4072, 4073, 4074, 4111, 4097, 1054, 1055, 1048, 1049, 1147, ODL25125-1342, ODL25125-1354, 1251, ODL25255-4530, ODL25255-4531, 4126, 4127, 4169, 4171, 4341, 4324, ODL25255-4536, ODL25255-4535, ODL25255-4397, 4141, 4143, 4144, 4146, ODL25255-4606, 4148, ODL25255-4577, ODL25255-4605, 4302, 4304, 4352, ODL25255-4352, 4330, ODL25255-4419, ODL25255-4534, ODL25255-4416, ODL25255-4401, 4168, 4107, 4255, 4257, 4252, 4256, 4254, 4253, 4236, ODL25255-4437, ODL25255-4443, ODL25255-4444, ODL25255-4438, 4294, 4293, ODL25255-4556, ODL25255-4382, ODL25255-4378, ODL25255-4377, ODL25255-4379, ODL25255-4380, ODL25255-4381, ODL25255-4420, ODL25255-4421
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1151-2018 | Class II | Terminated | Voluntary: Firm initiated |