Z-1151-2018Class II

AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125...

Siemens Healthcare Diagnostics Inc. / initiated 2018-01-18 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system on . FDA classified it as Class II and gives the reason as "A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1151-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.
Root cause tag
design
FDA root cause
Device Design
Action
A recall notification letter was sent to affected US consignees on 1/19/2018 notifying them of the issue. In addition, the recalling firm is instructing field service engineers to inspect the tip present sensor for misalignment at the next service visit.
Quantity in commerce
73 units in US
Distribution
Worldwide distribution. US nationwide, Algeria, Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
Product codes and lots
All units in the AutoMate 2500 Family are affected; Serial Numbers - 1107, 4048, 4055, 4025, 4052, 4079, 4091, 4076, 4072, 4073, 4074, 4111, 4097, 1054, 1055, 1048, 1049, 1147, ODL25125-1342, ODL25125-1354, 1251, ODL25255-4530, ODL25255-4531, 4126, 4127, 4169, 4171, 4341, 4324, ODL25255-4536, ODL25255-4535, ODL25255-4397, 4141, 4143, 4144, 4146, ODL25255-4606, 4148, ODL25255-4577, ODL25255-4605, 4302, 4304, 4352, ODL25255-4352, 4330, ODL25255-4419, ODL25255-4534, ODL25255-4416, ODL25255-4401, 4168, 4107, 4255, 4257, 4252, 4256, 4254, 4253, 4236, ODL25255-4437, ODL25255-4443, ODL25255-4444, ODL25255-4438, 4294, 4293, ODL25255-4556, ODL25255-4382, ODL25255-4378, ODL25255-4377, ODL25255-4379, ODL25255-4380, ODL25255-4381, ODL25255-4420, ODL25255-4421

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1151-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1151-2018Class IITerminatedVoluntary: Firm initiated