Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography...
Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system on . FDA classified it as Class II and gives the reason as "One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-0522-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- Philips Healthcare is notifying the affected users of this issue via this Customer Information Letter and correcting via Field Change Order FC088200514.
- Quantity in commerce
- 2
- Distribution
- One domestic and 1 foreign consignee ( Spain)
- Product codes and lots
- Ingenuity TF PET/CT Model # 882442
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0522-2018 | Class II | Terminated | Voluntary: Firm initiated |