Z-1169-2018Class II

AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number...

Siemens Healthcare Diagnostics Inc. / initiated 2018-01-19 / Completed / root cause: design

Beckman Coulter Inc. began a recall of AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255 on . FDA classified it as Class II and gives the reason as "Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination". As of , the recall status on record is Completed, and MDM tags the root cause as design.

Recall record

Recall number
Z-1169-2018
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.
Root cause tag
design
FDA root cause
Device Design
Action
Beckman Coulter plans to replace the belt via Service Modification during the next planned service maintenance visit. At that time, the customer will be informed of the belt replacement by the service engineer.
Quantity in commerce
98 units in the US
Distribution
Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States
Product codes and lots
All systems up to the following Serial Numbers: ODL25120 Revision T; SN3101, ODL25125 Revision W; SN1508, ODL25250 Revision T; SN2416, ODL25255 Revision W; SN4617

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1169-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1169-2018Class IICompletedVoluntary: Firm initiated