Hologic Panther(R) Fusion(TM), Catalog number ASY-09600
Hologic, Inc began a recall of Hologic Panther(R) Fusion(TM), Catalog number ASY-09600 Product Usage: The Panther instrument is a fully automated sample-to-result instrument, eliminating the... on . FDA classified it as Class II and gives the reason as "The Panther or the Panther Fusion glass panel separates from the 19 inch touch screen monitor and could potentially fall and hit an operator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0595-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hologic, Inc.
- Product
- Hologic Panther(R) Fusion(TM), Catalog number ASY-09600 Product Usage: The Panther instrument is a fully automated sample-to-result instrument, eliminating the need for batch processing and automates all aspects of nucleic acid testing on a single, integrated platform. The Panther can be test samples with random access or batch managed; can perform multiple assays on the same patient sample with up to 4 different assays simultaneously.
- Product code
- NSU Instrumentation For Clinical Multiplex Test Systems
- Reason for recall
- The Panther or the Panther Fusion glass panel separates from the 19 inch touch screen monitor and could potentially fall and hit an operator.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm sent an Urgent Medical Device Recall Notification dated January 8, 2018 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed that a Hologic representative will contact your site to arrange a visit in order to provide an interim solution to secure the glass panel until a replacement monitor with the issue corrected can be installed. For questions contact Hologic Technical Support. at +1-888-484-4747 or +1-858-410-8511 or by e-mail at [email protected]
- Quantity in commerce
- 103 units
- Distribution
- Worldwide distribution
- Product codes and lots
- Serial Numbers: 2090002307, 2090002308, 2090002309, 2090002310, 2090002311, 2090002312, 2090002339, 2090002340, 2090002341, 2090002342, 2090002343, 2090002344, 2090002345, 2090002346, 2090002347, 2090002348, 2090002349, 2090002350, 2090002351, 2090002352, 2090002353, 2090002359, 2090002360, 2090002361, 2090002362, 2090002363, 2090002366, 2090002367, 2090002368, 2090002369, 2090002370, 2090002371, 2090002372, 2090002373, 2090002374, 2090002381, 2090002382, 2090002383, 2090002384, 2090002385, 2090002387, 2090002388, 2090002389, 2090002390, 2090002391, 2090002393, 2090002404, 2090002405, 2090002406, 2090002407, 2090002409, 2090002411, 2090002413, 2090002414, 2090002415, 2090002420, 2090002421, 2090002422, 2090002428, 2090002429, 2090002430, 2090002432, 2090002433, 2090002434, 2090002435, 2090002436, 2090002437, 2090002438, 2090002446, 2090002447, 2090002448, 2090002449, 2090002450, 2090002451, 2090002452, 2090002458, 2090002462, 2090002465, 2090002466, 2090002467, 2090002468, 2090002469, 2090002470, 2090002471, 2090002472, 2090002473, 2090002476, 2090002481, 2090002483, 2090002484, 2090002485, 2090002338, 2090002394, 2090002417, 2090002431, 2090002463, 2090002464, 2090002492, 2090002491, 2090002490, 2090002489, 2090002488
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0595-2018 | Class II | Terminated | Voluntary: Firm initiated |