Z-0595-2018Class II

Hologic Panther(R) Fusion(TM), Catalog number ASY-09600

Hologic, Inc. / initiated 2018-01-20 / Terminated / root cause: design

Hologic, Inc began a recall of Hologic Panther(R) Fusion(TM), Catalog number ASY-09600 Product Usage: The Panther instrument is a fully automated sample-to-result instrument, eliminating the... on . FDA classified it as Class II and gives the reason as "The Panther or the Panther Fusion glass panel separates from the 19 inch touch screen monitor and could potentially fall and hit an operator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0595-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hologic, Inc.
Product
Hologic Panther(R) Fusion(TM), Catalog number ASY-09600 Product Usage: The Panther instrument is a fully automated sample-to-result instrument, eliminating the need for batch processing and automates all aspects of nucleic acid testing on a single, integrated platform. The Panther can be test samples with random access or batch managed; can perform multiple assays on the same patient sample with up to 4 different assays simultaneously.
Product code
NSU Instrumentation For Clinical Multiplex Test Systems
Reason for recall
The Panther or the Panther Fusion glass panel separates from the 19 inch touch screen monitor and could potentially fall and hit an operator.
Root cause tag
design
FDA root cause
Device Design
Action
The firm sent an Urgent Medical Device Recall Notification dated January 8, 2018 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed that a Hologic representative will contact your site to arrange a visit in order to provide an interim solution to secure the glass panel until a replacement monitor with the issue corrected can be installed. For questions contact Hologic Technical Support. at +1-888-484-4747 or +1-858-410-8511 or by e-mail at [email protected]
Quantity in commerce
103 units
Distribution
Worldwide distribution
Product codes and lots
Serial Numbers: 2090002307, 2090002308, 2090002309, 2090002310, 2090002311, 2090002312, 2090002339, 2090002340, 2090002341, 2090002342, 2090002343, 2090002344, 2090002345, 2090002346, 2090002347, 2090002348, 2090002349, 2090002350, 2090002351, 2090002352, 2090002353, 2090002359, 2090002360, 2090002361, 2090002362, 2090002363, 2090002366, 2090002367, 2090002368, 2090002369, 2090002370, 2090002371, 2090002372, 2090002373, 2090002374, 2090002381, 2090002382, 2090002383, 2090002384, 2090002385, 2090002387, 2090002388, 2090002389, 2090002390, 2090002391, 2090002393, 2090002404, 2090002405, 2090002406, 2090002407, 2090002409, 2090002411, 2090002413, 2090002414, 2090002415, 2090002420, 2090002421, 2090002422, 2090002428, 2090002429, 2090002430, 2090002432, 2090002433, 2090002434, 2090002435, 2090002436, 2090002437, 2090002438, 2090002446, 2090002447, 2090002448, 2090002449, 2090002450, 2090002451, 2090002452, 2090002458, 2090002462, 2090002465, 2090002466, 2090002467, 2090002468, 2090002469, 2090002470, 2090002471, 2090002472, 2090002473, 2090002476, 2090002481, 2090002483, 2090002484, 2090002485, 2090002338, 2090002394, 2090002417, 2090002431, 2090002463, 2090002464, 2090002492, 2090002491, 2090002490, 2090002489, 2090002488

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0595-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0595-2018Class IITerminatedVoluntary: Firm initiated