Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0669-2018Class II

LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for...

Synthes GmbH / initiated 2017-08-07 / Terminated / root cause: manufacturing process

Synthes, Inc. began a recall of LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments on . FDA classified it as Class II and gives the reason as "May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0669-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments
Product code
FZX Guide, Surgical, Instrument
Reason for recall
May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
In August of 2017 customers were mailed an URGENT FIELD SAFETY NOTICE titled "884115 - Locking Compression Plate (LCP) Drill Sleeve" requesting that customers who received product subject to recall to please contact: [email protected] or call 610-719-5450. In the event that you have product requiring use: Pre-assemble the drill sleeve with the plate preoperatively. If you experience difficulty connecting the drill sleeve to the plate, use the additional 1.1 mm locking drill sleeve provided in the graphic case, or: o Align the 1.1 mm drill bit to the plate as perpendicular as possible and then drill freehand and the locking screw can be engaged in this manner. o If the drill hole must be used and no option to lock remains, a non-locking screw can be used.
Quantity in commerce
567 units
Distribution
Distribution to US, Canada, and Germany.
Product codes and lots
Lot numbers FT00227 9972971 H028400 H161606 H288454

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0669-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0669-2018Class IITerminatedVoluntary: Firm initiated