Z-0974-2018Class II

CMI (Collagen Meniscus Implant) device, Ivy Sports Medicine s collagen-based meniscus implant

Stryker Corp. / initiated 2018-01-25 / Open, Classified / root cause: packaging

Stryker Corporation began a recall of CMI (Collagen Meniscus Implant) device, Ivy Sports Medicine s collagen-based meniscus implant Product Usage: The CMI device, Ivy Sports Medicine s... on . FDA classified it as Class II and gives the reason as "The recalled products were shipped without the required temperature control packaging, therefore could potentially have been exposed to elevated temperatures during transit". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0974-2018
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Stryker Corp.
Product
CMI (Collagen Meniscus Implant) device, Ivy Sports Medicine s collagen-based meniscus implant Product Usage: The CMI device, Ivy Sports Medicine s collagen-based meniscus implant, is comprised primarily of bovine type I collagen (nominally 99%) derived from tendon and small quantities of glycosaminoglycans (GAGs: chondroitin sulfate and sodium hyaluronate). The device functions as a resorbable scaffold that is replaced by the patient s own tissue.The CMI device is designed to function as an absorbable template to facilitate host meniscus tissue regeneration in patients who have an irreparable meniscus tear or loss of meniscus tissue. The CMI meniscus tissue through the implant s absorption and replacement by patient s native tissue.
Product code
OLC Scaffold, Partial Medial Meniscal Defects Extending Into The Red/White Zone, Resorbable Bovine Collagen
Reason for recall
The recalled products were shipped without the required temperature control packaging, therefore could potentially have been exposed to elevated temperatures during transit.
Root cause tag
packaging
FDA root cause
Packaging
Action
The international Stryker site was notified by email on 1/25/2018. Customers are instructed to: 1. Inform individuals within your organization who need to be aware of this device removal. 2. Review Part Numbers (4600, 4601, 4607, and 4612) for affected lot numbers. Please determine if you have the affected product in stock. Response is required. 3. If no product is found, notify your local Stryker office.4. If you do have product, segregate the product and call your local Stryker office to arrange for product return and issuance of credit.
Quantity in commerce
155 devices
Distribution
Netherlands
Product codes and lots
Serial/Lot Number(s) Affected: 4600: 15J456; 4601: 16A463, 16G470, 17F510; 4607: 16A461, 16D465, 16E467, 16F468, 17E506, 17E507; 4612: 16A458, 16A459, 17D489, 17E503, 17E504

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0974-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0974-2018Class IIOngoingVoluntary: Firm initiated