Z-0778-2019Class II

Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00 Indications for Use: The Haemonetics Cell...

Haemonetics Corporation / initiated 2017-08-02 / Terminated / root cause: manufacturing process

Haemonetics Corporation began a recall of Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00 Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components... on . FDA classified it as Class II and gives the reason as "Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0778-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Haemonetics Corporation
Product
Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00 Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
Product code
CAC Apparatus, Autotransfusion
Reason for recall
Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On August 2, 2017, the firm, Haemonetics, sent a customer notification letter titled "URGENT Medical Device Recall for 70ml Centrifuge Bowls" to all customers. The letter informs the customers of the risk to health, actions to be taken and the product and distribution information. Action to be Taken by Customer: Please discontinue use of the products listed in the letter and return any unused product to Haemonetics. Contact your local customer service representative at 800-537-2802 to coordinate the return, shipment and credit receipt of unused product. Customers were also instructed to complete and return the RECALL ACKNOWLEDGEMENT FORM BY FAX TO +1-781-356-3558 OR SCAN AND E-MAIL TO [email protected]. Customers with questions or concerns are instructed to call the Customer Support Team at 800-537-2802 or email at [email protected].
Quantity in commerce
11,208 units
Distribution
US Nationwide Distribution to states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, ME, MN, MO, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WA, and WI.
Product codes and lots
Catalog number: 0291A-00, 0291E-00

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0778-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0778-2019Class IITerminatedVoluntary: Firm initiated
Z-0779-2019Class IITerminatedVoluntary: Firm initiated