Liquid Cardiac Control CQ5053
Randox Laboratories, Limited began a recall of Liquid Cardiac Control CQ5053 on . FDA classified it as Class II and gives the reason as "Randox has confirmed that Troponin T and Troponin I in Liquid Cardiac quality control CQ5053 lot 4069CK does not meet the shelf-life claim in the product labelling. As the control may now fall outside the provided...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0937-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- Liquid Cardiac Control CQ5053
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Randox has confirmed that Troponin T and Troponin I in Liquid Cardiac quality control CQ5053 lot 4069CK does not meet the shelf-life claim in the product labelling. As the control may now fall outside the provided control ranges the product is to be recalled from the field.
- Root cause tag
- labeling and UDI
- FDA root cause
- Unknown/Undetermined by firm
- Action
- A recall notice was issued to customers on August 1, 2017 instructing them to discontinue use of product, discard remaining product for reimbursement and provide documentation of destruction.
- Quantity in commerce
- 450 kits
- Distribution
- Nationally
- Product codes and lots
- Lot 4069CK
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0937-2018 | Class II | Terminated | Voluntary: Firm initiated |