Z-0508-2018Class II

AArtis Q/ Q.Zen biplane Artis zee biplane

Siemens Medical Solutions USA, Inc. / initiated 2017-08-24 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging... on . FDA classified it as Class II and gives the reason as "Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0508-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Siemens sent an Urgent: Medical Device Correction letter dated August 24, 2017 to all affected customers. The letter identified the affected product, problem and actions to be taken. A local management or dispatch will schedule for an inspection to correct the issue. For questions please call 1-800-888-7436
Quantity in commerce
265 units
Distribution
US Nationwide Distribution
Product codes and lots
10094141 10848282 10848355

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0508-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0508-2018Class IITerminatedVoluntary: Firm initiated