AArtis Q/ Q.Zen biplane Artis zee biplane
Siemens Medical Solutions USA, Inc began a recall of AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging... on . FDA classified it as Class II and gives the reason as "Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0508-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Siemens sent an Urgent: Medical Device Correction letter dated August 24, 2017 to all affected customers. The letter identified the affected product, problem and actions to be taken. A local management or dispatch will schedule for an inspection to correct the issue. For questions please call 1-800-888-7436
- Quantity in commerce
- 265 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- 10094141 10848282 10848355
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0508-2018 | Class II | Terminated | Voluntary: Firm initiated |