BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used...
Becton Dickinson & Company began a recall of BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations... on . FDA classified it as Class II and gives the reason as "A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0968-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- In August of 2017 All US direct consignees were contacted via e-mail and/or UPS mail. An Urgent Product Removal Recall letter was issued, identifying the affected device and the reason for the recall. It also requested customers to review their inventory and quarantine any affected devices, which will be returned to the firm for replacement. The Customer Recall Response Form should be completed and returned regardless of whether there is any affected devices on hand. For questions or concerns please contact Customer Service at 1-855-215-4992.
- Quantity in commerce
- 8,494,000
- Distribution
- Nationwide Distribution
- Product codes and lots
- Catalog No. 367841 Lot No. 6279849 UDI: (01)30382903678410
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0968-2018 | Class II | Terminated | Voluntary: Firm initiated |