Z-0968-2018Class II

BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used...

Becton Dickinson & Co / initiated 2017-08-29 / Terminated / root cause: manufacturing process

Becton Dickinson & Company began a recall of BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations... on . FDA classified it as Class II and gives the reason as "A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0968-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
In August of 2017 All US direct consignees were contacted via e-mail and/or UPS mail. An Urgent Product Removal Recall letter was issued, identifying the affected device and the reason for the recall. It also requested customers to review their inventory and quarantine any affected devices, which will be returned to the firm for replacement. The Customer Recall Response Form should be completed and returned regardless of whether there is any affected devices on hand. For questions or concerns please contact Customer Service at 1-855-215-4992.
Quantity in commerce
8,494,000
Distribution
Nationwide Distribution
Product codes and lots
Catalog No. 367841 Lot No. 6279849 UDI: (01)30382903678410

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0968-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0968-2018Class IITerminatedVoluntary: Firm initiated