Z-0617-2018Class II
Norian Drillable Inject
Synthes (USA) Products LLC / initiated 2017-08-10 / Terminated / root cause: manufacturing process
Synthes USA began a recall of Norian Drillable Inject on . FDA classified it as Class II and gives the reason as "The contract supplier notified DePuy Synthes on June 16, 2017 that the kit expiration dates do not align with the earliest expiration date of the component lots within the kit. Some kit expiration dates were later than...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0617-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Norian Drillable Inject
- Product code
- OIS Calcium Salt Bone Void Filler, Drillable, Non-Screw Augmentation
- Reason for recall
- The contract supplier notified DePuy Synthes on June 16, 2017 that the kit expiration dates do not align with the earliest expiration date of the component lots within the kit. Some kit expiration dates were later than the earliest expiration date of the component lots. The expiration dates of some components are earlier than that listed on the kit.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On August 15, 2017 an Urgent Medical Device Recall notice, titled "895204 NORIAN DRILLABLE INJECT" was issued to customers instructing them to review inventory and quarantine all affected product subject to the recall. For questions. concerns or to obtain a return authorization Call 1-800-479-6329. Returned product will be credited.
- Quantity in commerce
- 239 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Product Number: 07.704.003S Lot Number: DSD9110
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0617-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0618-2018 | Class II | Terminated | Voluntary: Firm initiated |