Z-0495-2018Class II

Syngo.plaza Picture Archiving and Communication System (PACS) with software version VB20A; Model numbers...

Siemens Medical Solutions USA, Inc. / initiated 2017-08-18 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza Picture Archiving and Communication System (PACS) with software version VB20A; Model numbers: 10863171, 10863172, 10863173 Syngo.plaza is a Picture... on . FDA classified it as Class II and gives the reason as "Software update to correct several issues that include (1)Potential data loss, (2) Study mix-up, (3) Incorrect measurements on multi-frame images, (4) Dearchiving issue, and (5) Unauthorized access of data due to...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0495-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo.plaza Picture Archiving and Communication System (PACS) with software version VB20A; Model numbers: 10863171, 10863172, 10863173 Syngo.plaza is a Picture Archiving and Communication System (PACS) software device intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Software update to correct several issues that include (1)Potential data loss, (2) Study mix-up, (3) Incorrect measurements on multi-frame images, (4) Dearchiving issue, and (5) Unauthorized access of data due to inadequate permissions for shared folders.
Root cause tag
software
FDA root cause
Software design
Action
Siemens released a software update VB20A_HF04 and a Customer Information Letter was sent with Update Instruction SY010/17/P. Service reports for the affected sites will be monitored to ensure the recall was effective. For further questions, please call (610) 219-6461.
Quantity in commerce
47
Distribution
USA (nationwide) Distribution to the states of : AL, CT, FL, KS, LA, MA, MI, MN, MO, MS, NE, NY, OH, OR, PA, TX and WI. Foreign consignees were not reported as device is manufactured outside the U.S.
Product codes and lots
Syngo.plaza with software version VB20A; Model numbers: 10863171, 10863172, 10863173

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0495-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0495-2018Class IITerminatedVoluntary: Firm initiated