BC Thrombin Reagent Kit Lot Number 46751; UDI - 008427680131024675120180908 and BC Thrombin Reagent Kit Lot...
Siemens Healthcare Diagnostics began a recall of BC Thrombin Reagent Kit Lot Number 46751; UDI - 008427680131024675120180908 and BC Thrombin Reagent Kit Lot Number 47184; UDI -008427680131024718420190221; BC... on . FDA classified it as Class II and gives the reason as "The BC Thrombin Reagent kit lot 46751 (contains Thrombin Reagent lot 517468) and kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged Thrombin Time (TT) results for expected normal samples...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0827-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- BC Thrombin Reagent Kit Lot Number 46751; UDI - 008427680131024675120180908 and BC Thrombin Reagent Kit Lot Number 47184; UDI -008427680131024718420190221; BC Thrombin Reagent; Test, Thrombin Time Product Usage: For in vitro diagnostic use. Reagent for the quantitative determination of the thrombin time in citrated human plasma.
- Product code
- GJA Test, Thrombin Time
- Reason for recall
- The BC Thrombin Reagent kit lot 46751 (contains Thrombin Reagent lot 517468) and kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged Thrombin Time (TT) results for expected normal samples, and may recover above the upper limit of normal (< 21 sec.) as stated within the IFU (Instruction For Use).
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens sent an Urgent Field Safety Notice letter dated October 2017 to affected customer. The letter identified the affected product, problem and actions to be taken by the customer. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter and the Urgent Field Corrective Action letter within 30 days. For questions contact your Siemens Customers Care Center or local Siemens technical support representative.
- Quantity in commerce
- 7631 units
- Distribution
- Worldwide distribution. US nationwide, Austria, Belgium, Bosnia-Herzegovina, Croatia, Finland, France (incl. Guadeloupe, Martinique, Reunion, French Guinea overseas departments), Germany, Greece, Hungary, Italy, Kazakhstan, Macedonia, Latvia, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, Sweden, Turkey, Argentina, Brazil, El Salvador, Mexico, Peru, Uruguay, Canada, and Australia.
- Product codes and lots
- Manufactured: BC Thrombin Reagent Kit Lot 46751: 3673 kits BC Thrombin Reagent Kit Lot 47184: 3958 kits Total kits Distributed to Customers Domestic and Foreign: BC Thrombin Reagent Kit Lot 46751: 3549 kits BC Thrombin Reagent Kit Lot 47184: 2580 kits Product Code/ Lot# Date Manufactured Date Expired 46751 2016-11-01 2018-09-08 47184 2017-03-20 2019-02-21
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0827-2018 | Class II | Terminated | Voluntary: Firm initiated |