Heliomolar Refill 20x0.25g 140/A2, Product code 541502AN, Catalog number 541502
Ivoclar Vivadent, Inc. began a recall of Heliomolar Refill 20x0.25g 140/A2, Product code 541502AN, Catalog number 541502 on . FDA classified it as Class III and gives the reason as "Complaints were filed claiming the material consistency was too soft and too sticky in comparison to its usual consistency, causing unexpected handling for the user during packing and contouring". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0666-2018
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ivoclar Vivadent, Inc.
- Product
- Heliomolar Refill 20x0.25g 140/A2, Product code 541502AN, Catalog number 541502
- Product code
- EBF Material, Tooth Shade, Resin
- Reason for recall
- Complaints were filed claiming the material consistency was too soft and too sticky in comparison to its usual consistency, causing unexpected handling for the user during packing and contouring.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm sent letters to consignees on August 3, 2017 instructing them to segregate the recalled product and arrange for a replacement of stock.
- Quantity in commerce
- 3,255 units
- Distribution
- US and Canada
- Product codes and lots
- Batch: V37176, Exp. date: 15.10.2020; Batch: V40818, Exp. date: 13.11.2020; Batch: V45145, Exp. date: 05.12.2020; Batch: V51367, Exp. date: 22.01.2021; Batch: W00520, Exp. date: 29.01.2021; Batch: W01514, Exp. date: 04.02.2021; Batch: W07036, Exp. date: 05.03.2021; Batch: W13323, Exp. date: 07.04.2021
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0665-2018 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0666-2018 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0667-2018 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0668-2018 | Class III | Terminated | Voluntary: Firm initiated |