PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek...
Bard Access Systems Inc. began a recall of PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather... on . FDA classified it as Class II and gives the reason as "Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0536-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.
- Product code
- LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
- Reason for recall
- Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Customers were notified via letter on approximately 12/28/2017. Instructions include a reminder that if additional extension sets are attached to the luers of the catheters, they should be appropriately secured as per hospital recommended securement protocol, to refrain from having unsupported extension sets that could put stress on the luer extension leg junction of the catheter, and to notify customers if the product was further distributed. Also, the notification contained a statement that Bard Access is in the process of updating their Instructions for Use to include a statement regarding the importance of properly securing additional extension sets per hospital recommended securement protocol. On 04/11/2018, the recall was expanded and customers were notified via letter, which explained that the previous recall has been expanded to involve return of affected product as well as include product distributed from February 2017 through March 2018. Instructions included to examine inventory for affected product, quarantine and immediately discontinue distribution of the product, complete and return the Business Recall Response Form, arrange for the return of affected product, and notify customers if the product has been further distributed. For distributors, Bard Access Systems requested a customer list so that they could notify their customers.
- Quantity in commerce
- 53,699 units total
- Distribution
- Distribution US nationwide.
- Product codes and lots
- Product Code (Lot/Serial No.): S1274108D (REBN2334), S9274108D (REBN2336), S3274108PD (REBP1471), S3274335 (REBP1580), S3274355 (REBP1581), S3274355P (REBP1582), S3274335P (REBP1583), S1274108PD (REBP1971), S1274108 (REBQ1883), S3274355 (REBQ1884), S3274335 (REBQ1885), S3274108D (REBQ1886), S1274108D (REBQ1887), S9274108PD (REBQ1888), S0274108PD (REBQ1889), S3274335P (REBQ1890), S1274118 (REBQ1891), S9274108D (REBQ1892), S1274108D (REBR0160), S1274108D (REBN1883), S1173108D (REBN2329), S9173108D (REBN2332), S1274108PD (REBN2378), S9274108PD (REBN2379), S1173108PD (REBP0948), S3173335 (REBP0949), S3173355 (REBP0950), S9173108PD (REBP1020), s3173355p (REBP1023), S3173335P (REBP1024), S3274108D (REBQ1276), S1173108 (REBQ1882), S1173108 (REBQ2473), S1173108D (REBQ2474), S9173108D (REBQ2475), S1173108PD (REBQ2479), S0173108PD (REBR1399), S1173108PD (REBS0167), S3274108D (REBQ2476), S3173108PD (REBQ2480), S3173355 (REBQ2586), S3173335 (REBQ2587), s3173355p (REBQ2590), s3173108d (REBQ2592), S1173108PD (REBR1398), S0173108PD (REBS2081), S1274108D , REBS0062), S1274108D (REBS1035), S1274108D (REBS1567), S1274108D (REBR1347), S9274108D (REBQ2477), S3274355P (REBS0026), S3274335P (REBS0534), S3274335 (REBS0574), S3274355P (REBS1557), S3274355P (REBS2011), S3274355P (REBS2550), S1274108D (REBS2562), S1274118 (REBT0078), s1274108pd (REBT0146), S1274108PD (REBT0146), S3274355P (REBT0168), S3274355P (REBT0204), CK000765 (REBT0255), S1274108 (REBT0682), S0274108PD (REBT0697), S9274108D (REBT0754), S1274108D (REBT0755), S1274108PD (REBT0757), S3274335 (REBT0805), CK000765 (REBT0843), S9274108PD (REBT1446), S9274108 (REBT1885), CK000767 (REBT1889), S3274108PD (REBT2364), S9274108D (REBU0059), CK000767 (REBW1216), S3274355 (REBR1259), S3173355 (REBR1261), S1274108PD (REBR1345), S3274108PD (REBR1401), S1274108D (REBS0781), S3274355P (REBS0843), S1274108D (REBS1033), S1274108D (REBS2070), CK000759 (REBS2133), S3173355P (REBT0101), S3173335P (REBT0116), S3173355P (REBT0737), S3274335 (REBT1585), S1274108D (REBT1602), S3274355 (REBT1802), S3274108 (REBT1884), ck000770 (REBU0664), S9274108D (REBU0691), S1173108 (REBU1584), S9274108D (REBU1639), S1173108D (REBV0536), S1274108D (REBW0268), S3173108D (REBR1014), S1173108D (REBS1988), S3173108D (REBT0147), S9173108PD (REBT0234), S1173108D (REBT0240), S1173108D (REBT1398), S3173355P (REBT1801), S3173108 (REBT1882), S9173108 (REBT1883), S3173355P (REBT1948), S3173355P (REBU0082), S1173108D (REBU0293), S1173108D (REBU0690), S3173355P (REBU0714), S0173108PD (REBU1224), S3173355P (REBU1265), s9173108pd (REBU2255), S3173335P (REBV0052), S3173118 (REBV1102), S9173118 (REBV1103), S1173108PD (REBW1843), S3173108PD (REBX1664), s1173108 (REBT0688), S1274108D (REBU0296), S1274108D (REBU0765), S1274108 (REBU1047), S1274108D (REBU1271), S9173108D (REBU1640), S1274108PD (REBW0728), s3173355 (REBW1532), S1173108 (REBX1560), S1173108PD (REBX1663), S3173118 (REBY0008), S1173108D (REBY0782), S1173108PD (REBY0832), S0173108PD (REBY1316), S1173108 (REBY1472), S1173108D (REBY2516), CK000283A (REBV1203), S3173108PD (REBS0171), S3274108PD (REBT0233), S9274108PD (REBT1340), S3274108PD (REBT1403), S1274108 (REBT1846), s1274108 (REBT2073), S9274118 (REBU0001), S9274118 (REBU0002), S3274355 (REBU0083), S3274108D (REBU0297), S9274108PD (REBU0694), S3274355 (REBU0734), S3274355 (REBV0533), S3274118 (REBV1104), CK000114A (REBV2151), S1274108PD (REBV2327), S1274108D (REBW1840), S1274108D (REBX0325), CK000770 (REBX0460), S1274108D (REBX1581), CK000759 (REBX2038), S1274108D (REBX2436), S1274108D (REBX0326), S1274108PD (REBY0780), S1274108D (REBY0835), S1274108 (REBY1004), S1274108D (REBZ0814), CK000114A (REBZ1268), S1274108D (REBV1690), CK000770 (REBV2233), S1274108D (REBV2326), S1274108D (REBW0782), CK000759 (REBW1228), S1274108D (REBW1841), S1274108D (REBW1988), CK000767 (REBX0441), CK000759 (REBX0457), S1274108PD (REBX1122), S1274108D (REBX1240), S1274108PD (REBY1239), S1173108D (REBZ0958), S1173108PD (REBZ1627), S1274108D (RECN0085), S1274108PD (RECN0939), S1274108 (RECN1068), CK000114A (RECQ0041), S3173335 (REBY1941), S1173108D (REBY1982), S9173108PD (REBZ0816), S1173108D (REBZ0838), S1173108PD (REBZ0840), S3173108PD (REBZ0960), S1173108PD (RECN0088), S1173108D , RECN1199), s1173108pd (RECN1200), S3173108PD (RECN1406), S1173108D (RECN2175), S1173108PD (RECN2177), S0173108PD (RECN2678), S1173108D (RECP0357), S1173108PD (RECP0594), S1173108 (RECP1895), S3173355 (RECP2311), S1274108D (REBY0518), S3274108D (REBY0834), S1274108PD (REBY2518), S1173108PD (REBY2740), S9274108PD (REBZ0815), s1274108pd (REBZ1718), S1274108PD , RECN0027), S1274108PD (RECN2563), S3274108D (REBX2246), S3274335 (REBY0879), S1274108D (REBY2013), S3274335 (RECN0888), S1274108D (RECN0938), S1274108D (RECN2573), S3274108D (RECN2679), S9274108D (RECP0358), S1274108D (RECP0590), S3274355 (RECP0846), S1274108PD (RECP2482), S3274108D (RECQ0417), S3274108D (REBY2517), S1274108D (REBY2739), S3274355 (REBZ0494), S3274335 (REBZ0681), S3274108D (REBZ0839), S1274108D (REBZ0885), S1274108D (REBZ1625), S1274108 (REBZ1678), S1274108D (RECN1381), S1274108D (RECN2461), S3274108D (RECP0591), S9274108PD (RECP1511), S3274355 (REBZ1826), S3274108D (RECN0023), S9274108D (RECN0084), S9274108D (RECN1403), S1173108 (REBY1170)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0536-2018 | Class II | Terminated | Voluntary: Firm initiated |