510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0536-2018Class II

PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek...

Bard Access Systems, Inc. / initiated 2017-12-28 / Terminated / root cause: labeling and UDI

Bard Access Systems Inc. began a recall of PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather... on . FDA classified it as Class II and gives the reason as "Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0536-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.
Product code
LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Reason for recall
Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Customers were notified via letter on approximately 12/28/2017. Instructions include a reminder that if additional extension sets are attached to the luers of the catheters, they should be appropriately secured as per hospital recommended securement protocol, to refrain from having unsupported extension sets that could put stress on the luer extension leg junction of the catheter, and to notify customers if the product was further distributed. Also, the notification contained a statement that Bard Access is in the process of updating their Instructions for Use to include a statement regarding the importance of properly securing additional extension sets per hospital recommended securement protocol. On 04/11/2018, the recall was expanded and customers were notified via letter, which explained that the previous recall has been expanded to involve return of affected product as well as include product distributed from February 2017 through March 2018. Instructions included to examine inventory for affected product, quarantine and immediately discontinue distribution of the product, complete and return the Business Recall Response Form, arrange for the return of affected product, and notify customers if the product has been further distributed. For distributors, Bard Access Systems requested a customer list so that they could notify their customers.
Quantity in commerce
53,699 units total
Distribution
Distribution US nationwide.
Product codes and lots
Product Code (Lot/Serial No.): S1274108D (REBN2334), S9274108D (REBN2336), S3274108PD (REBP1471), S3274335 (REBP1580), S3274355 (REBP1581), S3274355P (REBP1582), S3274335P (REBP1583), S1274108PD (REBP1971), S1274108 (REBQ1883), S3274355 (REBQ1884), S3274335 (REBQ1885), S3274108D (REBQ1886), S1274108D (REBQ1887), S9274108PD (REBQ1888), S0274108PD (REBQ1889), S3274335P (REBQ1890), S1274118 (REBQ1891), S9274108D (REBQ1892), S1274108D (REBR0160), S1274108D (REBN1883), S1173108D (REBN2329), S9173108D (REBN2332), S1274108PD (REBN2378), S9274108PD (REBN2379), S1173108PD (REBP0948), S3173335 (REBP0949), S3173355 (REBP0950), S9173108PD (REBP1020), s3173355p (REBP1023), S3173335P (REBP1024), S3274108D (REBQ1276), S1173108 (REBQ1882), S1173108 (REBQ2473), S1173108D (REBQ2474), S9173108D (REBQ2475), S1173108PD (REBQ2479), S0173108PD (REBR1399), S1173108PD (REBS0167), S3274108D (REBQ2476), S3173108PD (REBQ2480), S3173355 (REBQ2586), S3173335 (REBQ2587), s3173355p (REBQ2590), s3173108d (REBQ2592), S1173108PD (REBR1398), S0173108PD (REBS2081), S1274108D , REBS0062), S1274108D (REBS1035), S1274108D (REBS1567), S1274108D (REBR1347), S9274108D (REBQ2477), S3274355P (REBS0026), S3274335P (REBS0534), S3274335 (REBS0574), S3274355P (REBS1557), S3274355P (REBS2011), S3274355P (REBS2550), S1274108D (REBS2562), S1274118 (REBT0078), s1274108pd (REBT0146), S1274108PD (REBT0146), S3274355P (REBT0168), S3274355P (REBT0204), CK000765 (REBT0255), S1274108 (REBT0682), S0274108PD (REBT0697), S9274108D (REBT0754), S1274108D (REBT0755), S1274108PD (REBT0757), S3274335 (REBT0805), CK000765 (REBT0843), S9274108PD (REBT1446), S9274108 (REBT1885), CK000767 (REBT1889), S3274108PD (REBT2364), S9274108D (REBU0059), CK000767 (REBW1216), S3274355 (REBR1259), S3173355 (REBR1261), S1274108PD (REBR1345), S3274108PD (REBR1401), S1274108D (REBS0781), S3274355P (REBS0843), S1274108D (REBS1033), S1274108D (REBS2070), CK000759 (REBS2133), S3173355P (REBT0101), S3173335P (REBT0116), S3173355P (REBT0737), S3274335 (REBT1585), S1274108D (REBT1602), S3274355 (REBT1802), S3274108 (REBT1884), ck000770 (REBU0664), S9274108D (REBU0691), S1173108 (REBU1584), S9274108D (REBU1639), S1173108D (REBV0536), S1274108D (REBW0268), S3173108D (REBR1014), S1173108D (REBS1988), S3173108D (REBT0147), S9173108PD (REBT0234), S1173108D (REBT0240), S1173108D (REBT1398), S3173355P (REBT1801), S3173108 (REBT1882), S9173108 (REBT1883), S3173355P (REBT1948), S3173355P (REBU0082), S1173108D (REBU0293), S1173108D (REBU0690), S3173355P (REBU0714), S0173108PD (REBU1224), S3173355P (REBU1265), s9173108pd (REBU2255), S3173335P (REBV0052), S3173118 (REBV1102), S9173118 (REBV1103), S1173108PD (REBW1843), S3173108PD (REBX1664), s1173108 (REBT0688), S1274108D (REBU0296), S1274108D (REBU0765), S1274108 (REBU1047), S1274108D (REBU1271), S9173108D (REBU1640), S1274108PD (REBW0728), s3173355 (REBW1532), S1173108 (REBX1560), S1173108PD (REBX1663), S3173118 (REBY0008), S1173108D (REBY0782), S1173108PD (REBY0832), S0173108PD (REBY1316), S1173108 (REBY1472), S1173108D (REBY2516), CK000283A (REBV1203), S3173108PD (REBS0171), S3274108PD (REBT0233), S9274108PD (REBT1340), S3274108PD (REBT1403), S1274108 (REBT1846), s1274108 (REBT2073), S9274118 (REBU0001), S9274118 (REBU0002), S3274355 (REBU0083), S3274108D (REBU0297), S9274108PD (REBU0694), S3274355 (REBU0734), S3274355 (REBV0533), S3274118 (REBV1104), CK000114A (REBV2151), S1274108PD (REBV2327), S1274108D (REBW1840), S1274108D (REBX0325), CK000770 (REBX0460), S1274108D (REBX1581), CK000759 (REBX2038), S1274108D (REBX2436), S1274108D (REBX0326), S1274108PD (REBY0780), S1274108D (REBY0835), S1274108 (REBY1004), S1274108D (REBZ0814), CK000114A (REBZ1268), S1274108D (REBV1690), CK000770 (REBV2233), S1274108D (REBV2326), S1274108D (REBW0782), CK000759 (REBW1228), S1274108D (REBW1841), S1274108D (REBW1988), CK000767 (REBX0441), CK000759 (REBX0457), S1274108PD (REBX1122), S1274108D (REBX1240), S1274108PD (REBY1239), S1173108D (REBZ0958), S1173108PD (REBZ1627), S1274108D (RECN0085), S1274108PD (RECN0939), S1274108 (RECN1068), CK000114A (RECQ0041), S3173335 (REBY1941), S1173108D (REBY1982), S9173108PD (REBZ0816), S1173108D (REBZ0838), S1173108PD (REBZ0840), S3173108PD (REBZ0960), S1173108PD (RECN0088), S1173108D , RECN1199), s1173108pd (RECN1200), S3173108PD (RECN1406), S1173108D (RECN2175), S1173108PD (RECN2177), S0173108PD (RECN2678), S1173108D (RECP0357), S1173108PD (RECP0594), S1173108 (RECP1895), S3173355 (RECP2311), S1274108D (REBY0518), S3274108D (REBY0834), S1274108PD (REBY2518), S1173108PD (REBY2740), S9274108PD (REBZ0815), s1274108pd (REBZ1718), S1274108PD , RECN0027), S1274108PD (RECN2563), S3274108D (REBX2246), S3274335 (REBY0879), S1274108D (REBY2013), S3274335 (RECN0888), S1274108D (RECN0938), S1274108D (RECN2573), S3274108D (RECN2679), S9274108D (RECP0358), S1274108D (RECP0590), S3274355 (RECP0846), S1274108PD (RECP2482), S3274108D (RECQ0417), S3274108D (REBY2517), S1274108D (REBY2739), S3274355 (REBZ0494), S3274335 (REBZ0681), S3274108D (REBZ0839), S1274108D (REBZ0885), S1274108D (REBZ1625), S1274108 (REBZ1678), S1274108D (RECN1381), S1274108D (RECN2461), S3274108D (RECP0591), S9274108PD (RECP1511), S3274355 (REBZ1826), S3274108D (RECN0023), S9274108D (RECN0084), S9274108D (RECN1403), S1173108 (REBY1170)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0536-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0536-2018Class IITerminatedVoluntary: Firm initiated