510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0848-2018Class II

Concorde Lift Driver Shaft AO. Used with supplemental internal spinal fixation systems that have been cleared...

Depuy Orthopaedics Inc. / initiated 2017-07-26 / Terminated

DePuy Orthopaedics, Inc. began a recall of Concorde Lift Driver Shaft AO. Used with supplemental internal spinal fixation systems that have been cleared for use in the lumbar spine on . FDA classified it as Class II and gives the reason as "Potential for Intra-operative breakage of driver tips". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0848-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Concorde Lift Driver Shaft AO. Used with supplemental internal spinal fixation systems that have been cleared for use in the lumbar spine.
Product code
MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Reason for recall
Potential for Intra-operative breakage of driver tips
FDA root cause
Under Investigation by firm
Action
On August 1, 2017, an Urgent Product Recall notice titled "CONCORDE LIFT DRIVER Driver Shaft and Torque Handle" was mailed to customers that received the affected instruments. The letter described the issue, potential hazard, and actions to be taken. The notice instructs customers to cease further distribution or use and to contact a DePuy Synthes Spine sales consultant to return the products subject to recall. Customers are to review, complete, sign, and return the business reply form provided to the firm within 5 business days of receipt of the notification. A copy of the notice should be forwarded to all staff that need to be informed, as well as, any facility that the affected device was further distributed to. Customers should direct any questions regarding this recall to their DePuy Synthes Spine Sales Consultant or the Clarke Madigan, DePuy Synthes Spine Recall Coordinator, at 508-828-609 or [email protected].
Quantity in commerce
130 units total
Distribution
Nationwide Distribution
Product codes and lots
Product code: 287804101G Lot number: 052316-A R, 121416-A R

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0848-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0848-2018Class IITerminatedVoluntary: Firm initiated
Z-0849-2018Class IITerminatedVoluntary: Firm initiated