Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use...
Roche Diagnostics Corporation began a recall of Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use, Catalog Numbers: 03739040692, 03023109001... on . FDA classified it as Class II and gives the reason as "Complaints have been received concerning questionable results on analyzers due to customers utilizing sample 13 mm tubes without tube adapters". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1269-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use, Catalog Numbers: 03739040692, 03023109001, 05023599001, 05023572001, 03739040001 & 03023109973.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Complaints have been received concerning questionable results on analyzers due to customers utilizing sample 13 mm tubes without tube adapters.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- Roche Diagnostics Corporation issued Urgent Medical Device Correction (UMDC) notices and Fax Forms to their customers dated 12/18/2017 via UPS Ground (receipt required). Customers are encouraged to do the following Mandatory Steps to improve the vertical alignment and handling of 13 mm sample tubes and false bottom tubes: FOR cobas e 411 & cobas e 601 analyzers and MODULAR ANALYTICS E 170 module: 1) Always use Roche rack cup adapters with 13 mm tubes 2) Ensure sample tubes are straight FOR cobas e 411 disk system: 1) Always use 13 mm Sample Disk Tube Adapters (SDTAs) with 13 mm sample tubes. 2) Ensure sample tubes are straight. 3) Ensure 13 mm SDTAs are not exposed to direct sunlight. 4) If a larger diameter tube is left inserted in a 13 mm SDTA for more than 3 days, the adapter may become distorted. Do not insert smaller diameter tubes into distorted adapter as they may not align correctly. **All customers are encouraged to DISCARD Analyzer Bulletin 10-009U3. Follow the care and handling instructions for using adapters as outlined in this Urgent Medical Device Correction (UMDC) and complete & return the UMDC fax form via fax to 1-877-271-1472
- Distribution
- US Nationwide
- Product codes and lots
- None
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1268-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1269-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1270-2018 | Class II | Terminated | Voluntary: Firm initiated |