Z-1270-2018Class II

cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers...

Roche Diagnostics GmbH / initiated 2017-12-18 / Terminated / root cause: labeling and UDI

Roche Diagnostics Corporation began a recall of cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack system), 04775279001 (disk system)... on . FDA classified it as Class II and gives the reason as "Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1270-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack system), 04775279001 (disk system), 04775201973 (rack system) & 04775279973 (disk system).
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
Roche Diagnostics Corporation issued Urgent Medical Device Correction (UMDC) notices and Fax Forms to their customers dated 12/18/2017 via UPS Ground (receipt required). Customers are encouraged to do the following Mandatory Steps to improve the vertical alignment and handling of 13 mm sample tubes and false bottom tubes: FOR cobas e 411 & cobas e 601 analyzers and MODULAR ANALYTICS E 170 module: 1) Always use Roche rack cup adapters with 13 mm tubes 2) Ensure sample tubes are straight FOR cobas e 411 disk system: 1) Always use 13 mm Sample Disk Tube Adapters (SDTAs) with 13 mm sample tubes. 2) Ensure sample tubes are straight. 3) Ensure 13 mm SDTAs are not exposed to direct sunlight. 4) If a larger diameter tube is left inserted in a 13 mm SDTA for more than 3 days, the adapter may become distorted. Do not insert smaller diameter tubes into distorted adapter as they may not align correctly. **All customers are encouraged to DISCARD Analyzer Bulletin 10-009U3. Follow the care and handling instructions for using adapters as outlined in this Urgent Medical Device Correction (UMDC) and complete & return the UMDC fax form via fax to 1-877-271-1472
Distribution
US Nationwide
Product codes and lots
None

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1270-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1268-2018Class IITerminatedVoluntary: Firm initiated
Z-1269-2018Class IITerminatedVoluntary: Firm initiated
Z-1270-2018Class IITerminatedVoluntary: Firm initiated