510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0970-2018Class III

Immunoassay Premium Plus Controls (3x4x5mL and 12x5mL packages)

Randox Laboratories, Limited / initiated 2017-08-11 / Terminated / root cause: design

Randox Laboratories began a recall of Immunoassay Premium Plus Controls (3x4x5mL and 12x5mL packages) on . FDA classified it as Class III and gives the reason as "The ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0970-2018
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Immunoassay Premium Plus Controls (3x4x5mL and 12x5mL packages)
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.
Root cause tag
design
FDA root cause
Device Design
Action
On July 24, 2017 an Urgent notice was issued instructing customers to place a copy a notice in all kits remaining in stock. This notifies customers that they should amend the claims for open stability in the instructions for use, to state use immediately after the 30 minute reconstitution period. Questions or concerns can be emailed to [email protected]
Quantity in commerce
199 Kits
Distribution
Distribution US nationwide.
Product codes and lots
Catalog numbers:  IA3109 (Lots: 1476EC, 1573EC, 1619EC) IA3110 (Lots 1477EDC, 1599EDC, 1621EC, 1662EC) IA3111 (Lots: 1478EC, 1508EC, 1578EC, 1647EC) IA3112 (Lots: 359199, 377412, 373302, 382147, 382679, 395668, 416979)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0970-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0969-2018Class IIITerminatedVoluntary: Firm initiated