Z-0662-2018Class II

Fuji Computed Radiography Mammography Suite, FCRMS (for CR-IR363AWS) The Fuji Computed Radiography...

FUJIFILM Healthcare Americas Corporation / initiated 2017-08-04 / Terminated / root cause: software

Fujifilm Medical Systems U.S.A., Inc. began a recall of Fuji Computed Radiography Mammography Suite, FCRMS (for CR-IR363AWS) The Fuji Computed Radiography Mammography Suite (FCRMS) is a software device that, in... on . FDA classified it as Class II and gives the reason as "FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR1000AWS, FDR-2000AWS, FDR-3000AWS, and CRIR363AWS assign a unique ID number to...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0662-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Fuji Computed Radiography Mammography Suite, FCRMS (for CR-IR363AWS) The Fuji Computed Radiography Mammography Suite (FCRMS) is a software device that, in conjunction with a specified Fuji Computed Radiography system forms the Fuji Computed Radiography for mammography (FCRm) device. FCRm with a dedicated mammographic x-ray machine generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film based mammographic (SFM) systems. The mammographic images can be interpreted by a qualified physician using either hardcopy film or softcopy display at a workstation.
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR1000AWS, FDR-2000AWS, FDR-3000AWS, and CRIR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
Root cause tag
software
FDA root cause
Software design
Action
The firm, FUJIFILM Medical Systems U.S.A., Inc. (FMSU), sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated 7/28/2017 to its customers. The letter described the products, problem and actions to be taken. The customers were instructed to please read and follow the instructions in the "ACTIONS TO BE TAKEN BY CUSTOMER/USER" section. If you have an error, contact your local FUJIFILM office; and complete and return the Customer Feedback Form. FUJIFILM service personnel will contact all of the medical facilities where the applicable products have been installed to arrange for this correction and visit to take corrective measures.. If you have any questions about this matter, call Recall contact (203) 276-3445 or email: [email protected].
Quantity in commerce
71 units (68 units Domestic. 3 units Foreign) in total
Distribution
Worldwide Distribution.
Product codes and lots
363AWS. Released Software Versions: V5.1, V6.0. Not Released Software Versions: V5.0, V5.2, V6.1, V7.0.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0662-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0659-2018Class IITerminatedVoluntary: Firm initiated
Z-0660-2018Class IITerminatedVoluntary: Firm initiated
Z-0661-2018Class IITerminatedVoluntary: Firm initiated
Z-0662-2018Class IITerminatedVoluntary: Firm initiated