Z-0560-2018Class II

Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm, Material CA30L1

Medtronic, Inc. / initiated 2017-08-07 / Open, Classified / root cause: design

Covidien LLC began a recall of Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm, Material CA30L1 on . FDA classified it as Class II and gives the reason as "The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0560-2018
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm, Material CA30L1
Product code
NEY System, Ablation, Microwave And Accessories
Reason for recall
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Root cause tag
design
FDA root cause
Device Design
Action
All consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.
Distribution
Nationwide including PR, Canada, China
Product codes and lots
UDI 10884521189553, Lot codes: S5LG004PX,S6AG001X,S6BG002X,S6BG005X,S6BG016PX,S6CG003X,S6CG005X,S6CG017X,S6DG004X,S6EG001X,S6FG001X,S6FG006X,S6FG010X,S6KG002X,S6KG006PRX,S6KG006PX,S6LG004X,S6MG002X,S7CG001PCX,S7CG011PCX,S7CG015CX,S7EG001X,S7EG003X,S7FG011PX

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0560-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0556-2018Class IIOngoingVoluntary: Firm initiated
Z-0557-2018Class IIOngoingVoluntary: Firm initiated
Z-0558-2018Class IIOngoingVoluntary: Firm initiated
Z-0559-2018Class IIOngoingVoluntary: Firm initiated
Z-0560-2018Class IIOngoingVoluntary: Firm initiated
Z-0561-2018Class IIOngoingVoluntary: Firm initiated