Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2
Covidien LLC began a recall of Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2 on . FDA classified it as Class II and gives the reason as "The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0561-2018
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2
- Product code
- NEY System, Ablation, Microwave And Accessories
- Reason for recall
- The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- All consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.
- Distribution
- Nationwide including PR, Canada, China
- Product codes and lots
- UDI 10884521706606, Lot codes: S6MG007PX,S6MG010PX,S6MG016PX,S7AG002PX,S7BG005X,S7BG009X,S7DG004X,S7FG006X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0556-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0557-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0558-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0559-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0560-2018 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0561-2018 | Class II | Ongoing | Voluntary: Firm initiated |