Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref...
Stryker Sustainability Solutions began a recall of Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134301, Sterile, Rx. The Reprocessed 2515 NAV eco... on . FDA classified it as Class II and gives the reason as "An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0916-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Sustainability Solutions
- Product
- Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134301, Sterile, Rx. The Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. They are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed 2515 NAV eco Variable EP Catheters provide location information when used with compatible CARTO EP Navigation Systems version 2.3 or higher.
- Product code
- NLH Catheter, Recording, Electrode, Reprocessed
- Reason for recall
- An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Stryker sent an Urgent Medical Device Recall letter dated January 2, 2018, explains the reason for recall, risk to health, and instructs them to discontinue use of the affected products. The customer is to check their inventory for the affected serial numbers provided in the attachment and complete the recall effectiveness check form to indicate the amount of product in inventory. Return instructions are provided. For further questions, please call (888) 763-8803.
- Quantity in commerce
- 170 devices
- Distribution
- USA (nationwide) Distribution was made to medical facilities in AZ, CA, CT, FL, GA, IL, KS, LA, MN, MO, OH, PA, TX, VT, WA, and Wi. Foreign distribution was made to Canada.
- Product codes and lots
- Serial numbers 2451387, 2484264, 2484471, 2486823, 2487380, 2488350, 2490592, 2490605, 2492930, 2494064, 2494067, 2495242, 2495335, 2495337, 2495339, 2500403, 2500911, 2510524, 2518512, 2523848, 2523849, 2523851, 2525678, 2528071, 2534854, 2534856, 2608429, 2609031, 2609102, 2609105, 2609106, 2609107, 2609985, 2610794, 2610802, 2614613, 2614847, 2614909, 2615094, 2615098, 2615099, 2615319, 2615646, 2615712, 2615722, 2615725, 2615937, 2615938, 2615940, 2615943, 2615952, 2615966, 2616087, 2616088, 2616089, 2616090, 2616212, 2617794, 2617920, 2617987, 2617999, 2618010, 2618258, 2618832, 2618837, 2618881, 2618884, 2618990, 2618992, 2618993, 2619615, 2619742, 2619744, 2619827, 2620937, 2620942, 2620945, 2621283, 2621384, 2621385, 2621387, 2621424, 2621425, 2621847, 2621860, 2621880, 2623972, 2623974, 2624006, 2624010, 2624113, 2624138, 2625427, 2625536, 2625550, 2625623, 2625625, 2626330, 2626348, 2626929, 2626966, 2626988, 2627596, 2627946, 2627948, 2627949, 2629162, 2629164, 2630084, 2630085, 2630087, 2630351, 2631186, 2631188, 2631189, 2631191, 2631680, 2632641, 2633968, 2634008, 2634252, 2634271, 2634596, 2634709, 2634815, 2634836, 2635075, 2635077, 2635342, 2635343, 2635867, 2635873, 2636658, 2636759, 2636906, 2636986, 2637009, 2637022, 2637023, 2637233, 2637234, 2637237, 2637239, 2637257, 2637308, 2637495, 2637497, 2640103, 2640106, 2640107, 2640118, 2640119, 2642853, 2643222, 2643426, 2643432, 2646002, 2647122, 2669947, 2693960, 2696255, 2705129, 2708915, 2708916, 2712736, 2712756, 2713556, 2713593, 2713595, and 2713596.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0916-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0917-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY