Z-0717-2018Class II

Allura Xper FD and and Xper FD ORT - Philips X-ray Systems

Philips North America LLC / initiated 2017-12-20 / Terminated / root cause: design

Philips Electronics North America Corporation began a recall of Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd for: Cardiovascular and... on . FDA classified it as Class II and gives the reason as "Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0717-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper FD20, Allura Xper FD20/10, Allura Xper FD20/15 and Allura Xper FD10/10 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and ElectroPhysiology. Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room (OR Table)
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
Root cause tag
design
FDA root cause
Device Design
Action
Philips sent an Urgent Field Safety Notice letter December 20, 2017 dated to all affected customers. The letter identified the affected product, problem and the actions to be taken. For question contact your local Philips representative.
Quantity in commerce
1486
Distribution
Worldwide Distribution - US Nationwide
Product codes and lots
All Allura Xper FD and and Xper FD ORT manufactured from July 2003 - May 2017 with system Codes: 722001  722002  722003 722010 722026 722005 722011 722027 722006 722012 722028 722008 722013  722015 722023 722035 722020  722029  722038  722039  722058  722059

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0717-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0717-2018Class IITerminatedVoluntary: Firm initiated