510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0917-2018Class II

Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref...

Stryker Sustainability Solutions / initiated 2018-01-02 / Terminated / root cause: design

Stryker Sustainability Solutions began a recall of Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134302, Sterile, Rx. The Reprocessed 2515 NAV eco... on . FDA classified it as Class II and gives the reason as "An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0917-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Sustainability Solutions
Product
Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134302, Sterile, Rx. The Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. They are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed 2515 NAV eco Variable EP Catheters provide location information when used with compatible CARTO EP Navigation Systems version 2.3 or higher.
Product code
NLH Catheter, Recording, Electrode, Reprocessed
Reason for recall
An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker sent an Urgent Medical Device Recall letter dated January 2, 2018, explains the reason for recall, risk to health, and instructs them to discontinue use of the affected products. The customer is to check their inventory for the affected serial numbers provided in the attachment and complete the recall effectiveness check form to indicate the amount of product in inventory. Return instructions are provided. For further questions, please call (888) 763-8803.
Quantity in commerce
8 devices
Distribution
USA (nationwide) Distribution was made to medical facilities in AZ, CA, CT, FL, GA, IL, KS, LA, MN, MO, OH, PA, TX, VT, WA, and Wi. Foreign distribution was made to Canada.
Product codes and lots
Serial numbers 2468846, 2468849, 2509029, 2489074, 2491174, 2500529, 2465989, and 2500528

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0917-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0916-2018Class IITerminatedVoluntary: Firm initiated
Z-0917-2018Class IITerminatedVoluntary: Firm initiated