510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0627-2018Class II

AGC 2000 LP Patella, labeled as AGC Porous Patellar SM, Model Number 150802

Zimmer GmbH / initiated 2017-08-25 / Terminated

Zimmer Biomet, Inc. began a recall of AGC 2000 LP Patella, labeled as AGC Porous Patellar SM, Model Number 150802. Product Usage: For use in Total Knee Arthroplasty on . FDA classified it as Class II and gives the reason as "The affected AGC 2000 LP Patella products were over-molded leaving the outside diameter nonconforming to the print dimension". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0627-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
AGC 2000 LP Patella, labeled as AGC Porous Patellar SM, Model Number 150802. Product Usage: For use in Total Knee Arthroplasty.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The affected AGC 2000 LP Patella products were over-molded leaving the outside diameter nonconforming to the print dimension.
FDA root cause
Process change control
Action
Consignees were notified via mail on approximately 08/25/2017. Instructions to distributors, sales representatives, and distributor operations managers include to ensure affected team members are aware of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide any additional hospitals and/or surgeons that received direct shipments. Instructions to risk managers and surgeons include to ensure affected personnel are informed of the recall, quarantine all affected product, return any affected product in inventory, and complete and return the Certificate of Acknowledgment. Questions or concerns can be sent to the Zimmer Biomet customer call center at 1-574-371-3071 between 8:00 am and 5:00 pm EST, Monday through Friday. Or they can be sent via email to [email protected].
Quantity in commerce
28 devices
Distribution
US Nationwide Distribution to IN, MO, NY, TX.
Product codes and lots
Model No. 150802: Lot No. 305330 (UDI (01)00880304240230(17)220506(10)305330), 305340 ((01)00880304240230(17)220506(10)305330), 305350 ((01)00880304240230(17)220517(10)305350), 356750 ((01)00880304240230(17)220523(10)356750).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0627-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0627-2018Class IITerminatedVoluntary: Firm initiated
Z-0628-2018Class IITerminatedVoluntary: Firm initiated