Z-0706-2018Class II

Artis Q and Q.Zen, Model Numbers:10848280, 10848281, 10848282, 10848353 & 10848355

Siemens Medical Solutions USA, Inc. / initiated 2017-06-06 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Artis Q and Q.Zen, Model Numbers:10848280, 10848281, 10848282, 10848353 & 10848355 on . FDA classified it as Class II and gives the reason as "Potential for liquid entry". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0706-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis Q and Q.Zen, Model Numbers:10848280, 10848281, 10848282, 10848353 & 10848355
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Potential for liquid entry
Root cause tag
design
FDA root cause
Device Design
Action
On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail. Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461.
Distribution
U.S. Nationwide
Product codes and lots
10094135, 10094137, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848353, 10848354, 10848355, 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350, 7728392

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0706-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0706-2018Class IITerminatedVoluntary: Firm initiated
Z-0707-2018Class IITerminatedVoluntary: Firm initiated
Z-0708-2018Class IITerminatedVoluntary: Firm initiated